16 Channel 500 V/1 mA - 100µA Full Floating Channel Dual Range Board
The CAEN Mod. A1612 is a single width board (5TE wide) that houses 16 individual floating channels, delivered through DB37 connector. The output voltage is 0 ÷ 500 V with 10 mV set/monitor resolution. The maximum current value is 1 mA with a 100 nA setting resolution (low range: 100 μA; 10 nA setting resolution).
|
DB37 connector Single width (5 TE wide), 16 channels for Mod. A1612 |
Consult our connectors reference page for technical information.
The channels share a Individual Floating Channels, which allows on-detector grounding, reducing the noise level. A global enable/disable connector allows to disable the channels. The voltage ramp rates may be set independently for each channel and offer dual current ranges (software selectable).
Safety features allows the module to perform as a current generator and includes:
|
Channels |
can be enabled or disabled through the Global Interlock logic. |
|
Overvoltage and Undervoltage warning |
when the output voltage differs from the programmed value. |
|
Overcurrent detection |
when a channel attempts to exceed the programmed current limit, it signalled to be in “overcurrent” and enter in a TRIP status. The output voltage is varied to keep the current below the programmed limit for a programmable TRIP time, then the channel is switched off. If TRIP is set to “constant current mode”, the channel behaves like a current generator. |
|
Hardware VMAX |
maximum output voltage can be set, via front panel potentiometer, at the same common value for all the board channels. VMAX value can be read out via software. |
|
Safety Board Interlock |
this protection disables the HV generation when the HV outputs are not connected to their loads. |
16 independently controllable full floating channels with dual current range
500 V (1 mA / 100 µA) output ranges
1mV voltage monitor resolution
Current monitor resolution:
high range: 1 nA
low range: 100 pA
Programmable voltage ramp up / down rates
Typical voltage ripple smaller than 3 mVpp (1 KHz ÷ 20 MHz)
Overcurrent programmable protections
Programmable TRIP parameter
DB37 connector
Current generator operation in overcurrent condition