The CAEN Mod. A1520P houses 12 positive HV channels; each output channel has an independent floating ground. The output voltage can be programmed and monitored in the range 0 ÷ +500 V in 1 mV steps (17 bit resolution). All output channels are provided with Remote Sensing Lines to compensate for the voltage drop over the cables.
The board is provided with both current and voltage protections. For each channel, it is also possible to set an output current limit in the 0 ÷ 15 mA range, in 250 nA steps.
The output current is monitored with 25 nA resolution. If overcurrent occurs, the relevant channel can be programmed either to turn off after a programmable trip time or to keep on providing the maximum allowed current: this particular feature allows the module to work as current generator.
A global enable/disable connector allows to disable the channels when it is not terminated on 50 Ω. The HV Ramp-Up and Ramp-Down rates may be selected independently for each channel in the 1 ÷ 50 V/s range (1 V/s step). Moreover the floating channels allow on-detector grounding, thus allowing to reduce the noise level. The board features the AMP 75-pin output connector.
|
AMP 75-pin output connector 10TE-wide, 6U-high mechanics, 12 channels for Mod. A1520P |
Consult our connectors reference page for technical information.
12 independently controllable High Voltage channels
500 V / 15 mA output range
Positive polarity
Individual floating channels
AMP Multipin connector
17 bit voltage setting / monitor resolution
±1 ppm/°C thermal stability
Low ripple
Under/over-voltage alert, overcurrent and max. voltage protection
Interlock logic for unit enable
Software Tool for easy channel management